Certification Scope & Range

This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2017. This accreditation demonstrates technical competence for a defined scope and the operation of a laboratory quality management system (refer to joint ISO-ILAC-IAF Communiqué dated April 2017).

Scope of Accreditation to ISO/IEC 17025:2017 and ANSI/NCSL Z540-1-1994 (R2002)

Parameter / Equipment Range Uncertainty of Measurement (+/-) Reference Standard, Method
and/or Equipment
pH Meters (4, 7 and 10) pH 0.012 pH Standard Buffer Solutions
Parameter / Equipment Range Uncertainty of Measurement (+/-) Reference Standard, Method / Equipment
Parameter / Equipment Range Uncertainty of Measurement (+/-) Reference Standard, Method / Equipment
DC Voltage - Source Up to 330 mV
330 mV to 3.3 V
(3.3 to 33) V
(33 to 330) V
330 V to 1 kV
14 µV/V + 1.2 µV
6.3 µV/V + 10 µV
7.7 µV/V + 81 µV
12 µV/V + 0.78 mV
14 µV/V + 1.4 mV
Multifunction Calibrator
DC Voltage - Measure Up to 100 mV
100 mV to 1 V
(1 to 10) V
(10 to 100) V
100 V to 1 kV
9.6 µV/V + 1.1 µV
3.8 µV/V + 10 µV
8.4 µV/V + 3.3 µV
10 µV/V + 38 µV
10 µV/V + 0.13 mV
High Resolution DMM
Up to 6 kV
(6 to 20) kV
(20 to 35) kV
Up to 150 kV
10 mV/V + 0.6 V
20 mV/V + 2.4 V
90 mV/V + 51 V
5.1 mV/V + 6.9 V
DMM with High Voltage Probe
Thermocouple Simulation and Measure J-type thermocouple
(63 to 1 473) K
(-210 to 1 200) °C
0.24 K
(0.24 °C)
Multifunction Calibrator
K-type thermocouple
(73 to 1 645) K
(-200 to 1 372) °C
0.25 K
(0.25 °C)
S-type thermocouple
(273 to 1 673) K
(0 to 1 400) °C
0.52 K
(0.52 °C)
T-type thermocouple
(23 to 673) K
(-250 to 400) °C
0.25 K
(0.25 °C)
E-type thermocouple
(23 to 1 273) K
(-250 to 1 000) °C
0.43 K
(0.43 °C)
N-type thermocouple
(73 to 1 573) K
(-200 to 1 300) °C
0.37 K
(0.37 °C)
AC Voltage - Source (1 to 33) mV
(10 to 45) Hz
45 Hz to 10 kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 500) kHz
0.62 mV/V + 4.8 µV
0.12 mV/V + 4.7 µV
0.15 mV/V + 4.8 µV
0.78 mV/V + 4.7 µV
2.7 mV/V + 9.4 µV
6.2 mV/V + 39 µV
Multifunction Calibrator
(33 to 330) mV
(10 to 45) Hz
45 Hz to 10 kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 500) kHz
0.39 mV/V + 6.8 µV
0.11 mV/V + 7.1 µV
0.12 mV/V + 7.4 µV
0.27 mV/V + 6.7 µV
0.62 mV/V + 25 µV
1.6 mV/V + 54 µV
330 mV to 3.3 V
(10 to 45) Hz
45 Hz to 10 kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 500) kHz
0.23 mV/V + 43 µV
0.11 mV/V + 66 µV
0.14 mV/V + 58 µV
0.23 mV/V + 42µV
0.54 mV/V + 0.1 mV
1.9 mV/V + 0.47 mV
(3.3 to 33) V
(10 to 45) Hz
45 Hz to 10 kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(3.3 to 33) V
(10 to 45) Hz
45 Hz to 10 kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(33 to 330) V
45 Hz to 1 kHz
(1 to 10) kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
0.11 mV/V + 2.4 mV
0.15 mV/V + 5.5 mV
0.19 mV/V + 5.4 mV
0.23 mV/V + 5.3 mV
1.6 mV/V + 39 mV
330 V to 1 kV
45 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
0.23 mV/V + 8.1 mV
0.19 mV/V + 8 mV 0.23 mV/V + 8.2 mV
AC Voltage - Measure Up to 10 mV
(1 to 40) Hz
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
0.47 mV/V + 4 µV
0.14mV/V + 3 µV
0.22 mV/V + 3 µV
0.89 mV/V + 2.7 µV
High Resolution DMM, Multifunction Calibrator
(10 to 100) mV
(1 to 40) Hz
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
70 µV/V + 4.3 µV
70 µV/V + 2.1 µV
0.14 mV/V + 2.1 µV
0.3 mV/V + 2.1 µV
100 mV to 1 V
(1 to 40) Hz
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
70 µV/V + 40 µV
70 µV/V + 21 µV
0.14 mV/V + 21 µV
0.3 mV/V + 23 µV
(1 to 10) V
(1 to 40) Hz
40 H to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
70 µV/V + 0.4 mV
70 µV/V + 0.22 mV
0.14 mV/V + 0.21 mV
0.3 mV/V + 0.21 mV
(10 to 100) V
(1 to 40) Hz
40 H to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
0.2 mV/V + 4 mV
0.2 mV/V + 2 mV
0.2 mV/V + 2.1 mV
0.35 mV/V + 2.1 mV
100 V to 1 kV
(1 to 40) Hz
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
0.4 mV/V + 40 mV
0.4 mV/V + 20 mV
0.6 mV/V + 20 mV
1.1 mV/V + 79 mV
Up to 6 kV
60 Hz
10 mV/V + 5 V
(6 to 35) kV
60 Hz
51 mV/V + 9 V
Up to 150 kV
1 kHz
8.5 mV/V + 38 V
DC Current - Source Up to 330 µA
330 µA to 3.3 mA
(3.3 to 330) mA
(33 to 330) mA
330 mA to 1.1 A
(1.1 to 3) A
(3 to 11) A
59 µA/A + 61 nA
73 µA/A + 57 nA
77 µA/A + 0.21 µA
75 µA/A + 2.8 µA
0.16 mA/A + 31 µA
0.29 mA/A + 31 µA
0.42 mA/A + 31µA
Multifunction Calibrator
(10 to 16.5) A
(16.5 to 150) A
(150 to 1 000) A
4.7 mA/A + 29 mA
4.7 mA/A + 0.21 A
4.7 mA/A + 0.99 A
Multifunction Calibrator with Current Coil
DC Current - Measure Up to 100 nA
100 nA to 1 µA
(1 to 10) µA
(10 to 100) µA
100 µA to 1 mA
(1 to 10) mA
(10 to 100) mA
100 mA to 1 A
16 µA/A + 45 pA
11 µA/A + 54 pA
20 µA/A + 0.1 nA
20 µA/A + 0.81 nA
15 µA/A + 14 nA
20 µA/A + 51 nA
35 µA/A + 0.51 µA
35 µA/A + 5.5 µA/A
High Resolution DMM
AC Current - Source (29 to 330) µA
(10 to 20) Hz
(20 to 45) Hz
45 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(10 to 30) kHz
1.6 mA/A + 78 nA
1.2 mA/A + 78 nA
0.97 mA/A + 78 nA
2.3 mA/A + 0.12 µA
6.2 mA/A + 0.16 µA
12 mA/A + 0.31 µA
Multifunction Calibrator
330 µA to 3.3 mA
(10 to 20) Hz
(20 to 45) Hz
45 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(10 to 30) kHz
1.6 mA/A + 0.13 µA
0.97 mA/A + 0.12 µA
0.78 mA/A + 0.12 µA
1.6 mA/A + 0.16 µA
3.9 mA/A + 0.23 µA
7.8 mA/A + 0.47 µA
(3.3 to 33) mA
(10 to 20) Hz
(20 to 45) Hz
45 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(10 to 30) kHz
1.4 mA/A + 1.6 µA
0.7 mA/A + 1.6 µA
0.31 mA/A + 1.6 µA
0.62 mA/A + 1.6 µA
1.6 mA/A + 2.3 µA
3.1 mA/A + 3.1 µA
(33 to 330) mA
(10 to 20) Hz
(20 to 45) Hz
45 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(10 to 30) kHz
1.4 mA/A + 16 µA
0.7 mA/A + 16 µA
0.31 mA/A + 16 µA
0.78 mA/A + 39 µA
1.6 mA/A + 78 µA
3.1 mA/A + 0.16 mA
330 mA to 3 A
(10 to 45) Hz
45 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
1.4 mA/A + 78 µA
0.47 mA/A + 78 µA
4.7 mA/A + 0.78 mA
19 mA/A + 3.9 mA
(3 to 11) A
(45 to 100) Hz
100 Hz to 1 kHz
(5 to 10) kHz
0.45 A/A + 1.9 mA
0.77 A/A + 1.6 mA
23 mA/A + 1.6 mA
(10 to 16.5) A
(45 to 65) Hz
(65 to 440) Hz
5.5 mA/A + 33 mA
10 mA/A + 35 mA
Multifunction Calibrator with Current Coil
(16.5 to 150) A
(45 to 65) Hz
(65 to 440) Hz
5.6 mA/A + 0.27 A
10 mA/A + 0.27 A
(150 to 1 000) A
(45 to 65) Hz
(65 to 440) Hz
5.1 mA/A + 1.7 A
12 mA/A + 1.1 A
AC Current - Measure 100 µA to 1 mA
(20 to 45) Hz
(45 to 100) Hz
100 Hz to 5 kHz
1.5 mA/A + 0.2 µA
0.6 mA/A + 0.2 µA
0.3 mA/A + 0.2 µA
High Resolution DMM
(1 to 10) mA
(20 to 45) Hz
(45 to 100) Hz
100 Hz to 5 kHz
(5 to 20) kHz
4 mA/A + 2 µA
1.5 mA/A + 2 µA
0.6 mA/A + 2 µA
0.3 mA/A + 2 µA
(10 to 100) mA
(20 to 45) Hz
(45 to 100) Hz
100 Hz to 5 kHz
(5 to 20) kHz
4 mA/A + 20 µA
1.5 mA/A + 20 µA
0.6 mA/A + 20 µA
0.3 mA/A + 20 µA
AC Current - Measure 100 mA to 1 A
(20 to 45) Hz
(45 to 100) Hz
100 Hz to 5 kHz
(5 to 20) kHz
4 mA/A + 0.2 mA
1.6 mA/A + 0.2 mA
0.6 mA/A + 0.2 mA
1 mA/A + 0.2 mA
High Resolution DMM
Resistors - Source Fixed Values (at 1 kΩ) 24.9 Ω
375.6 Ω
5.97 kΩ
95.3 kΩ
6.9 mΩ
51 mΩ
0.79 Ω
12 Ω
Standard Resistors Kit
Multifunction Calibrator Up to 11 Ω
(11 to 33) Ω
(33 to110) Ω
(110 to 330) Ω
330 Ω to 1.1 kΩ
(1.1 to 3.3) kΩ
(3.3 to 11) kΩ
(11 to 33) kΩ
27 µΩ/Ω + 1.2 mΩ
19 µΩ/Ω + 1.7 mΩ
20 µΩ/Ω + 1.4 mΩ
21 µΩ/Ω + 2 m Ω
22 µΩ/Ω + 1.8 mΩ
20 µΩ/Ω + 22 mΩ
22 µΩ/Ω + 17 mΩ
19 µΩ/Ω + 0.3 Ω
Multifunction Calibrator
Resistance - Source (33 to 110) kΩ
(110 to 330) kΩ
330 kΩ to 1.1 MΩ
(1.1 to 3.3) MΩ
(3.3 to 11) MΩ
(11 to 33) MΩ
(33 to 110) MΩ
(110 to 330) MΩ
330 MΩ to 1.1 GΩ
19 µΩ/Ω + 0.52 Ω
24 µΩ/Ω + 2.7 Ω
24 µΩ/Ω + 3.8 Ω
24 µΩ/Ω + 0.1 kΩ
95 µΩ/Ω + 0.12 kΩ
0.17 mΩ/Ω + 3.1 kΩ
0.38 mΩ/Ω + 3.7 kΩ
2.3 mΩ/Ω + 81 kΩ
12 mΩ/Ω + 0.4 MΩ
Multifunction Calibrator
Resistance - Measure Up to 10 Ω
(10 to 100) Ω
100 Ω to 1 kΩ
(1 to 10) kΩ
(10 to 100) kΩ
100 kΩ to 1 MΩ
(1 to 10) MΩ
(10 to 100) MΩ
100 MΩ to 1 GΩ
14 µΩ/Ω + 75 µΩ
12 µΩ/Ω + 0.52 mΩ
10 µΩ/Ω + 0.57 mΩ
9.7 µΩ/Ω + 13 mΩ
10 µΩ/Ω + 57 mΩ
15 µΩ/Ω + 2.1 Ω
49 µΩ/Ω + 0.12 kΩ
0.16 mΩ/Ω + 79 kΩ
4.5 mΩ/Ω + 0.56 MΩ
High Resolution DMM
RTD Simulation Pt 385 (100 Ω)
(73 to 1 073) K
(-200 to 800) °C
0.09 K
(0.09 °C)
High Resolution DMM
385 (1 000 Ω)
(73 to 903) K
(-200 to 630) °C
0.1 K
(0.1 °C)
Pt 3916 (100 Ω)
(73 to 903) K
(-200 to 630) °C
0.09 K
(0.09 °C)
Pt 3926 (100 Ω)
(73 to 903) K
(-200 to 903) °C
0.09 K
(0.09 °C)
Ni 120 (120 Ω)
(193 to 533) K
(-80 to 260) °C
0.13 K
(0.13 °C)
Pt 385 (200 Ω)
(73 to 903) K
(-200 to 630) °C
0.1 K
(0.1 °C)
Pt 385 (500 Ω)
(73 to 903) K
(-200 to 630) °C
0.1 K
(0.1 °C)
Capacitance – Source
10 Hz to 1 kHz Charge/Discharge Rate (3.3 to 11) nF
(11 to 33) nF
(33 to 110) nF
(110 to 330) nF
1.8 mF/F + 11 pF
1.7 mF/F + 95 pF
1.7 mF/F + 0.13 nF
1.4 mF/F + 0.57 nF
Multifunction Calibrator
10 Hz to 600 Hz Charge/Discharge rate (0.33 to 1.1) µF 1.9 mF/F + 0.91 mF
10 Hz to 300 Hz Charge/Discharge rate (1.1 to 3.3) µF 1.4 mF/F + 5.7 nF
10 Hz to 150 Hz Charge/Discharge rate (3.3 to 11) µF 1.8 mF/F + 10 nF
10 Hz to 120 Hz Charge/Discharge rate (11 to 33) µF 2.6 mF/F + 52 nF
10 Hz to 80 Hz Charge/Discharge rate (33 to 110) µF 3.4 mF/F + 88 nF
Up to 50 Hz Charge/Discharge rate (110 to 330) µF 3 mF/F + 0.51 µF
Up to 50 Hz Charge/Discharge rate 330 µF to 1 mF 3.4 mF/F + 0.99 µF
Capacitance - Source Fixed Values 1 nF
1 kHz
0.28 nF Standard Capacitors
1 µF
100 Hz
120 Hz
1.5 nF
1.5 nF
1 kHz
10 µF
100 Hz
120 Hz
1 kHz
1.5 nF
15 nF
15 nF
15 nF
100 µF
100 Hz
120 Hz
1 kHz
0.15 µF
0.15 µF
0.15 µF
1 mF
100 Hz
120 Hz
1 kHz
1.9 µF
2.1 µF
2.1 µF
10 mF
100 Hz
120 Hz
1 kHz
0.11 mF
0.11 mF
0.15 mF
Inductance - Source Fixed Values 10 mH
100 Hz
1 kHz
6.5 µH
6.5 µH
Standard Inductors
Oscilloscopes Bandwidth (Leveled Sine Wave) 50 kHz to 600 MHz (4.8 + 0.006 8X) % Multifunction Calibrator
DC Voltage
50 Ω load
1 M Ω load
(0 to 6.6) V
(0 to 130) V
1.9 mV/V + 0.37 mV
0.46 mV/V + 0.5 mV
Square Wave -Amplitude
50 Ω load
1 M Ω load
(0 to 6.6) V
(0 to 130) V
0.19 mV/V + 0.46 mV
0.77 mV/V + 0.64 mV
Rise Time 3.5 ns Pulse Edge 41 ps
Oscilloscopes Time Marker (2 to 10) ns
(20 to 100) ns
(100 to 500) ns
(1 to 20) ms
(50 to 500) ms
(1 to 5) s
2.9 ns/s + 7.8 ps
27 ns/s + 7.7 ps
0.15 µs/s + 7.7 ps
4.6 ns/s + 8.6 ns
1.4 ns/s + 44 ns
2.8 ms/s + 9 ms
Multifunction Calibrator
Parameter / Equipment Range Uncertainty of Measurement (+/-) Reference Standard, Method / Equipment
Parameter / Equipment Range Uncertainty of Measurement (+/-) Reference Standard, Method / Equipment
RF Absolute Power
Generate Sine Wave into 50 Ω
200 Hz to 81 MHz
(-86.98 to 13.01) dBm 0.015 dBm Synthesized Level Generator
Sine Wave into 75 Ω
200 Hz to 81 MHz
(-88.74 to 11.25) dBm 0.015 dBm
RF Absolute Power - Generate
< 1.0 Hz to 100 KHz
(0.10 to 20) MHz
(-56 to 23) dBm 0.33 dBm Function Generator
RF Absolute Power - Generate 100 kHz to 2.060 GHz (-140 to 13) dBm 0.099 dBm Signal Generator
RF Absolute Power - Generate 10 MHz to 2.3 GHz (-9.95 to 10) dBm 0.018 Signal Generator
RF Absolute Power - Generate (2.3 to 26.5) GHz (-79.95 to -10) dBm 0.061 dBm Signal Generator
RF Absolute Power - Generate (20 to <26.5) GHz (-100 to -80) dBm 0.074 dB Signal Generator
RF Absolute Power Measure
10 MHz to 18 GHz
10 MHz to 18 GHz
(0 to 35) dBm
(35 to 44) dBm
0.12 dBm
0.23 dBm
RF Power Meter and Power Sensor
0.1 MHz to 4.2 GHz
0.1 MHz to 4.2 GHz
(-30 to 10) dBm
(10 to 20) dBm
0.08 dBm
0.09 dBm
10 MHz to 18 GHz
10 MHz to 18 GHz
(-70 to -30) dBm
(-30 to -20) dBm
0.1 dBm
0.11 dB
10 MHz to 18 GHz
10 MHz to 18 GHz
(-30 to 10) dBm
(10 to 20) dBm
0.08 dBm
0.17 dBm
30 MHz to 26.5 GHz
30 MHz to 26.5 GHz
(-20 to -10) dBm
(-10 to 30) dBm
0.15 dBm
0.15 dBm
Horn Antennas – Antenna Factors
Gain dBi Gain Numeric 700 MHz to 18 GHz
1-meter
3-meter
 
1.7 dB
0.9 dB
Signal Generator, Power Meter, Power Sensor, Spectrum Analyzer
Parameter / Equipment Range Uncertainty of Measurement (+/-) Reference Standard, Method / Equipment
Parameter / Equipment Range Uncertainty of Measurement (+/-) Reference Standard, Method / Equipment
Micrometers – Outside 3 Up to 4 in
(4 to 20) in
(20 to 36) in
(44 + 16L) µin
(32 + 22L) µin
(32 + 22L) µin
Gage Blocks, Optical Flats
Calipers - Outside Jaws 3 Up to 6 in
(6 to 40) in
(420 + 3.9L) µin
(350 + 17L) µin
Gage Blocks
Calipers - Inside Jaws 3 Up to 24 in
(24 to 40) in
(480 + 3.9L) µin
(340 + 19L) µin
Reference Bar, Gage Blocks
Calipers - Depth 3 Up to 24 in (530 + 1.7L) µin Gage Blocks, Surface Plate
Height Gages 3 Up to 24 in
(24 to 40) in
(490 + 10L) µin
(260 + 19L) µin
Reference Bar, Surface Plate, Test Indicator
Micrometers - Inside 3
(Head Movement Only)
Up to 1 in (81 + 24L) µin Gage Blocks, Gage Holder
Micrometers - Inside 3
(Resolution 0.0001 in)
 
(Resolution 0.001 in)
Up to 6 in
(6 to 24) in
(24 to 40) in
(100 + 12L) µin
(38 + 22L) µin
(460 + 17L) µin
Reference Bar, Gage Blocks
Micrometers - Depth 3 Up to 12 in (630 + 4.5L) µin Gage Blocks, Surface Plate
Bore Gages 3
(Resolution 0.000 1 in)
(0.1to 0.5) in
(0.5 to 3) in
(80 + 2L) µin
(150 + 19L) µin
Master Ring Gages
Indicators3
Test, Dial, Digital
(Resolution 0.000 1 in)
Up to 2 in (68 + 25L) µin Gage Blocks, Calibration Tester, Surface Plate
Flatness Up to 4 in 5.5 µin Master Flat
Optical Comparator 3
Horizontal Readout
Vertical Readout
Up to 8 in
Up to 8 in
(740 + 8.6L) µin
(760 + 8.7L) µin
Reading Scale
Thickness (Feeler) Gages (0 to 0.05) in 125 µin Digital Micrometer
Rulers 3 Up to 40 in (3 200 + 112L) µin Caliper
Plain Plugs 2 Up to 90 mm 4.3 µm + 0.002 8 µm/mm IAC Master Scanner
Plain Ring 2 2.5 to 100 mm 4.3 µm+ 0.004 7 µm/mm IAC Master Scanner
Thread Flank Angle
Measurements 2
Up to 60 ° 0.15 deg + 0.000 073 deg/deg IAC Master Scanner
Thread Plug Gages 2
Major / Minor Diameter Up to 90 mm 4.2 µm+ 0.016 µm/mm IAC Master Scanner
Effective Pitch Diameter Up to 90 mm 4.3 µm+ 0.016 µm/mm
Pitch 0.1 to 40 mm 1.4 µm + 0.003 7 µm/mm
Thread Ring Gages 2
Major / Minor Diameter 2.5 to 100 mm 4.9 µm + 0.012 µm/mm IAC Master Scanner
Effective Pitch Diameter 2.5 to 100 mm 5.1 µm + 0.01 µm/mm
Pitch 0.1 to 40 mm 1.6 µm+ 0.003 3 µm/mm
Parameter / Equipment Range Uncertainty of Measurement (+/-) Reference Standard, Method / Equipment
Parameter / Equipment Range Uncertainty of Measurement (+/-) Reference Standard, Method / Equipment
Pressure 2 - Hydraulic, Cross Floating 41.4 kPa to 16.5 MPa
(6 to 2 400) psi
0.22 kPa + 0.19 Pa/Pa
0.032 psi to 0.000 028 psi/psi
Comparison to Ruska 2400 Standard Dead Weight Tester
207 kPa to 82.7 MPa
(30 to 12 000) psi
0.35 kPa + 0.25 Pa/Pa
0.05 psi + 0.000 037 psi/psi
Up to 499.86 MPa
(Up to 72 500) psi
4.8 MPa + 0.000 072 MPa/MPa
700 psi + 0.000 000 5 psi/psi
Pressure Transducer
Pressure - Hydraulic Up to 499.86 MPa
(Up to 72 500) psi
4.8 MPa + 0.000 072 MPa/MPa
700 psi + 0.000 000 5 psi/psi
Pressure Transducer
(34.48 to 137.92) MPa
(5 000 to 20 000) psi
77 kPa + 0.000 55 Pa/kPa
11.2 psi + 0.000 08 psi/psi
Precision Pressure Monitor
Hardness Testers
Rockwell HRBW Scale
Low
Mid
High
2 HRBW
1.6 HRBW
1.8 HRBW
Indirect Comparison to Hardness Test Blocks
HRC Scale
Low
Mid
High
1.3 HRC
1.3 HRC
1.1 HRC
Rockwell Superficial HR30TW Scale
Low
Mid
High
1.6 HR30TW
1.3 HR30TW
1.4 HR30TW
Parameter / Equipment Range Uncertainty of Measurement (+/-) Reference Standard, Method / Equipment
Temperature at Ice Point 273.15 K
(0 °C)
0.033 K
0.033 °C
Standard Multimeter and Platinum Resistance Thermometer
Temperature - Measure (73 to 933) K
(-200 to 660) °C
0.000 004 K/K + 0.032 K
0.000 004 °C/°C + 0.032 °C
Standard Multimeter and Platinum Resistance Thermometer
(0 to 1 750) °C 0.004 3 °C/°C + 0.57 °C Type R Thermocouple and Multifunction Calibrator
Humidity - Source (10 to 95) %RH 0.81 %RH + 0.014 %RH/%RH Humidity Chamber and Humidity Meter
Humidity - Measure (10 to 95) %RH 0.7 %RH + 0.015 %RH/%RH
Parameter / Equipment Range Uncertainty of Measurement (+/-) Reference Standard, Method / Equipment
Frequency - Source Using Calibrator’s Normal Output (0.01 to 120) Hz
120 Hz to 1.2 kHz
(1.2 to 12) kHz
(12 to 120) kHz
120 kHz to 1.2 MHz
(1.2 to 2) MHz
1.2 µHz/Hz + 0.11 mHz
1.6 µHz/Hz + 0.48 mHz
1.9 µHz/Hz + 0.4 mHz
1.9 µHz/Hz + 1.2 mHz
1.9 µHz/Hz + 1.2 mHz
1.9 µHz/Hz + 14 mHz
Multifunction Calibrator
Frequency - Source Using Calibrator’s Oscilloscope Output 50 kHz to 100 MHz
(100 to 300) MHz
(300 to 600) MHz
1.9 µHz/Hz + 5.1 Hz
1.9 µHz/Hz + 0.3 Hz
1.9 µHz/Hz + 70 mHz
Frequency - Measure 1 Hz to 10 MHz 0.5 mHz/Hz + 0.1 µHz High Resolution DMM
Stopwatches Up to 24 hours 0.19 s NIST UTC Phone Time Signal
Frequency – Source and Measure 0.1 µHz to 3 GHz
10 Hz to 26.5 GHz
0.012 Hz + 0.385 Hz/MHz
0.17 Hz
Frequency Counter

Calibration and Measurement Capability (CMC) is expressed in terms of the measurement parameter, measurement range, expanded uncertainty of measurement and reference standard, method, and/or equipment. The expanded uncertainty of measurement is expressed as the standard uncertainty of the measurement multiplied by a coverage factor of 2 (k=2), corresponding to a confidence level of approximately 95%.

Notes:

  1. Mobile and On-site calibration service is available for most parameters, since on-site conditions are typically more variable than those in the laboratory, larger
    measurement uncertainties are expected on-site than what is reported on the accredited scope. Marsh Mobile Lab environmental conditions are controlled to meet the
    conditions needed to achieve the uncertainties listed in the Calibration and Measurement Capability (CMC) of the laboratory.
  2. These parameters can only be performed at the laboratory’s fixed location.
  3. X = measured value, F = measured value in lbf, L = Length in inches.
  4. This scope is formatted as part of a single document including Certificate of Accreditation No. AC-1172.